17. Error and system messages

Error messages for the measurement
Display
Meaning
Battery voltage weak, can only
function for a short period
With mains adapter operation:
incorrect voltage
Battery is depleted
With mains adapter operation:
incorrect voltage
No display
Battery is depleted
or
Confusing
With mains adapter operation:
symbols
incorrect voltage/poling
System error
Device does
not react
to pushing
Device defective
buttons
SENS
Sensor error: no sensor cable
ERRO
connected
Sensor breaks or device defective
ERR.1
Measurement range exceeded
Incorrect sensor connected
Sensor or device defective
ERR.2
Measurement range undercut
Incorrect sensor connected
Sensor, cable or device defective
ERR.7
System error
If "bAt" blinks in the display, the battery is depleted. Measurement can continue for a brief
time. If "bAt" remains in the display, the battery is fully depleted and must be replaced. No
more measurements can take place.
54
Remedy
Insert new battery
Check / replace mains adapter
Insert new battery
Check / replace mains adapter
Insert new battery
Check / replace mains adapter
Disconnect battery and mains
adapter, wait a short time and plug
in again
Send in for repair
Connect sensor
Send in for repair
Is measurement value above per-
missible range? -> Measurement
value is too high!
Check sensor
Send in for repair
Is measurement value below per-
missible range? -> Measurement
value is too low!
Check sensor
Send in for repair
Send in for repair
SD_310_2 07/2018