Data Logger; Manual Recording ("Func-Stor") - Lovibond Tintometer SD 310 Oxi Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 86

10. Data logger

Logger operation not possible with Auto-Range! A fixed preselection for the
measurement range must be made – see chapter 6 "Configuration of the
device" -
The device has two different logger functions:
"Func-Stor":
"Func-CYCL":
The logger records the conductivity and the temperature for each data set.
A data set is comprised of: Measurement cond/rES/TDS/SAL (one of these)
For the evaluation and transmission of the data, the software GSOFT3050 (V3.0 or later) is
required. It can be used to start and adjust the logger function very simply.
With the logger function activated (Func Stor of Func CYCL), the hold function is not available;
then the "store" button is only responsible for the logger operation.

10.1 Manual recording ("Func-Stor")

a) Record measurements manually:
If the "Func Stor" logger function was selected (see "Configuration of the device"),
up to 1000 measurements can be saved manually:
or
b) Call up manual recording:
Saved data sets can be read in the GSOFT3050 software and can be observed in the device
display itself.
i
or
48
Manual measurement recording at the push of the "store" button
In addition, a measurement entry (L-Id) is requested.
Automatic cyclical recording in the adjusted time interval
Temperature measurement value
L-ID measurement point (only with "Func-Stor")
Time and date at which the data is saved
Press briefly: Data set is saved ("St. XX" is briefly
displayed. XX is the number of the data set.)
measurement point input "L-Id": Selection of the measurement point
with keys.
Number from 0 - 19999.
The entry is confirmed.
If the logger memory is full, the following appears
Press and hold for two seconds:
The following appears in the display:
"rEAd LoGG" only appears if the data sets have already been
saved! Without data sets, the configuration menu
appears
Press briefly: Switch between measurement values, measurement point
and date+time display of the data set
Switch between the calibration data sets
Close the display of the records
SD_310_2 07/2018
Tabla de contenido
loading

Tabla de contenido