Lovibond Tintometer SD 310 Oxi Manual De Instrucciones página 39

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 86
Menu
Parameter
A
VTO
HLD
P.OFF
L,TE
0V T
ADR
DA.,N
DA[.0
DA[. 1
(*)
If there is data in the logger memory, parameters marked with (*) cannot be opened.
If they should be changed, the data must first be deleted!
SD 310_2 07/2018
OFF
Standard hold function at the push
of a button (only with Logger = oFF)
1 - 365
Time interval for calibration reminder
(in days)
OFF
No calibration memory
1 - 120
Auto power off (shut-off delay) in
minutes. If no button is pressed and
data exchange no longer takes place
over the interface, the device shuts off
automatically after the lapse of this time
(factory setting: 20 min)
OFF
Automatic shut-off deactivated (contin-
uous operation)
OFF
No background lighting, lowest
power consumption
5 - 120
Automatic shut-off of lighting after
5 - 120 s (factory setting: 5 s)
ON
Background lighting always on
OFF
No output function, lowest power
consumption
SER
Device output is serial interface
(factory setting)
DA[
Device output is analog output
0 1, 1 1,2 1, ...
Base address of the device for serial
9 1
interface communication (factory
setting 01)
[ONC
Analog output based on
concentration in ppm or mg/l
SAT or P 02
Analog output based on setting of
(H 2
0 . 0 - 1 0 0 . 0
Input of O2 concentration at which
%O2
the 0V analog output should be
output (factory setting 0.0 %O
0 . 0 - 1 0 0 . 0
Input of O
%O2
the 1V analog output should be
output (factory setting 100.0 %O
2
concentration at which
2
see
)
)
2
39
Tabla de contenido
loading

Tabla de contenido