This Method Is Adapted From Merck; Before Performing The Test Read The Original Test Instructions (Delivered With The Test) And; The Msds (Available At Www.merckmillipore.com); Spectroquant ® Is A Registered Trade Mark Of The Company Merck Kgaa - Lovibond Tintometer Manual Del Usuario

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 128
Notes:

1. This method is adapted from MERCK.

2. Before performing the test read the original test instructions (delivered with the test) and

the MSDS (available at www.merckmillipore.com).

3. Spectroquant
®
is a registered trade mark of the company MERCK KGaA.
4. Appropriate safety precautions and good lab technique should be used during the whole
procedure.
5. Because reaction depends on temperature, sample and tube temperature must be
between 20 and 25°C.

6. Sample volume should always be metered by using volumetric pipette (class A).

7. The test sample should have a pH value between 3 and 9.
8. Triton
is a registered trade mark of the company DOW Chemical Company.
®
9.
Triton
X-100
®
NP 10
Reagent / Accessories
MERCK Spectroquant
Nonylphenol Ethoxylat
1)
07/2017
Form of reagent/Quantity
1.01787.0001
Cell Test / 25 Tests
®
Order-No.
420764
57