Instrument Configuration - Lovibond Tintometer SD 310 Oxi Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 86
Menu
Parameter
SET
Set Configuration: General settings
U N,T
(ON F
(ONC
[H 2
RES
SAL
LCD.2
UN,T
T
[AL
[AL.P
[.INT
(*)
If there is data in the logger memory, parameters marked with (*) cannot be opened.
If they should be changed, the data must first be deleted!
38
Values
Meaning
(ONC
Display unit of oxygen concentration
ppm
ppm (same values as mg/l)
(ONC
Display unit of oxygen concentration
mg/l
mg/l (same values as ppm)
SAT
Upper display shows O
%O
2
(mg/l or ppm) or O
(factory setting)
P 02
Upper display shows O
hPa
or O
P 02
Upper display shows O
mmHg
or O
K
Best O
LO
Low O
display
OFF
Salinity correction deactivated (facto-
ry setting)
0.1 - 70.0
Salinity correction of oxygen
measurements, unit ‰ = PSU
T
Lower display always show temperature
P
Lower display always show absolute
pressure
P T
Lower display alternately shows abso-
lute pressure and temperature
°C
All temperatures specified in degrees
Celsius (factory setting)
°F
All temperatures specified in degrees
Fahrenheit
A,R
The oxygen sensor is calibrated in the
air or in gas (factory setting)
A Q UA
The sensor is calibrated in water
1-PT
1-point calibration in air (A,R) or in
air-saturated water (A Q UA)
2-PT / 3-PT
2 or 3-point calibration in air (A,R)/
air-saturated water (A Q UA), or in
oxygen/oxygen-saturated water or in
nitrogen/zero solution
1 - 365
Time interval for calibration reminder
(in days)
OFF
No calibration reminder
concentration
2
saturation in %
2
concentration
2
partial pressure in hPa
2
concentration
2
partial pressure in mmHg
2
display resolutions
2
display resolutions, more stable
2
SD_310_2 07/2018
see
*
Tabla de contenido
loading

Tabla de contenido