Operating Information; Principles Of Operation; System Setup - ResMed S8 Compact System Manual De Instrucciones

Ocultar thumbs Ver también para S8 Compact System:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

O
PERATING
P
O
RINCIPLES OF
PERATION
The
S8 C
OMPACT
obstructive sleep apnea (OSA). The
tubing to the mask worn by your patient. The continuous air stream 'splints' open the
upper airway, preventing airway collapse.
The
S8 C
OMPACT
beginning of treatment more comfortable, you may set a ramp so the pressure starts
low and gradually increases to full treatment pressure over the ramp time.
S
S
YSTEM
ETUP
1
S8 C
Place the
CAUTION
!
Be careful to place the device where it cannot be bumped, and where no one will
trip over the power cord.
Note: The S8 Compact device may be placed on the floor beside or underneath the
patient's bed. Ensure that the area is free from dust and clear of bedding, clothes, or any
other objects that could block the air inlet.
2
Connect the AC power cord to the corresponding socket on the rear of the
S8 C
(D-1). Plug the other end of the power cord into the appropriate power
OMPACT
outlet.
WARNING
!
Make sure the power cord and plug are in good condition and the equipment is
not damaged.
The air filter cover protects the device in the event of accidental liquid spillage
onto the device. Ensure an air filter and the air filter cover are fitted at all times.
3
Connect one end of the air tubing firmly onto the air outlet of the device (D-2) and
the other end to the mask system (D-3).
When the
S8 C
the unit is performing self-checks.
The
S8 C
OMPACT
CAUTION
!
When calibrating the S8 Compact or setting the treatment pressure, assemble the
device as the patient would normally use it, and connect all accessories that are
normally used.
I
NFORMATION
system is for clinical use and home treatment of adult patients with
delivers pressure within the range 4 to 20 cm H
on a table near the head of the patient's bed.
OMPACT
unit is first powered, the product name screen appears while
OMPACT
system is now ready for use (D-4).
S8 C
provides fixed-pressure room air via
OMPACT
O. To make the
2
O
I
PERATING
NFORMATION
11