nd
The AirCompact aerosol is intended for use in the electromagnetic environment specified below. The
client or user of the AirCompact aerosol must make sure that the device is used in this environment
Emissions test
Radiated/Conducted Emissions
CISPR11
Radiated/Conducted Emissions
42/
CISPR11
Harmonic Emissions IEC/EN
61000-3-2
Voltage Fluctuations/Flicker
Emissions IEC/EN 61000-3-3
The AirCompact aerosol is intended for use in the electromagnetic environment specified below. The
client or user of the AirCompact aerosol must make sure that the device is used in this environment
e-
Emissions test
m-
ed
Electrostatic discharge (ESD)
in
IEC/EN 61000-4-2
Electrical fast transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN61000-4-5
Voltage dips, short interruptions
and voltage variations IEC/EN
61000-4-11
Magnetic field IEC/EN 61000-
4-8
Immunity to radiation
disturbances IEC/EN 61000-4-6
Immunity to radiation
disturbances IEC/EN 61000-4-3
Note: U
is the mains voltage
T
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Group 1
Class B
Class A
[Satisfies]
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
±6 kV on contact
±8 kV in air
±2 kV Supply
±1 kV differential mode
<5% UT for 0.5 cycle
40% UT for 05 cycle
70% UT for 25 cycle
<5% UT
for 5 sec
3 A/m
3 Vrms 150 kHz to 80 MHz (for
non-life-supporting equipment)
3 V/m 80 MHz to 2.5 GHz (for
non-life-supporting equipment)
Compliance
The AirCompact aerosol uses RF
energy for its internal operation
only. Therefore its RF emissions
are very low and do not cause any
interference in nearby electronic
equipment.
The AirCompact aerosol is suitable
for use in all environments, includ-
ing domestic ones and those directly
connected to the public power net-
work that supplies buildings used
for domestic purposes.
Compliance
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should not ex-
ceed 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the AirCompact aerosol
requires continued operation, it
is recommended that the unit is
powered from an uninterruptible
power supply.
The magnetic field should be that
of a typical commercial or hospi-
tal environment.
1
Electromagnetic
Environment - Guidance
Electromagnetic
Environment - Guidance
GB