GB
suring that the lead connector remains at the
base of the compartment to enable the flap to
close. If the flap does not close, remove the
battery holder and replace it correctly.
• Re-close the flap.
• If the batteries are sufficiently charged,
proceed to therapy; otherwise, start the
recharge cycle.
• See table 1 for information on the bat-
teries' charge.
TABLE 1
LED light
Flashing red light
Flashing red light
Flashing red light
Fixed red light
No light on
No light on
No light on
LED light
Green flashing light
Green
flashing light (alarm)
Fixed green light
No light on
No light on
Status
Device operating with
rechargeable batteries
Device operating with
alkaline batteries
Device blocked
Device blocked
Device blocked
Device blocked
Device blocked
Phase
While charging
and
orange
While charging
While charging
At start of recharge cycle
While charging
Operation with alkaline batteries – ONLY
IF ABSOLUTELY NECESSARY
Use AA 1.5 V alkaline batteries only if ab-
solutely necessary.
In this case it is essential to ensure that
you NEVER START THE RECHARGE CYCLE
when alkaline batteries are inserted.
Never mix alkaline and rechargeable
batteries in the battery holder. Never re-
charge alkaline batteries.
Problem/Cause
Low battery charge
Start recharge cycle as soon
as possible
None (LED lights up after 4
Use alkaline batteries until
minutes max. when operat-
flat, then replace
ing with alkaline batteries)
Overheated motor
Wait for device to cool (Red LED
turns off) before restarting
Short circuit overload
Seek technical assistance
Faulty supply or internal
Seek technical assistance
technical problem
Polarity inversion
Unplug from power supply,
check that batteries are in-
serted correctly and restart.
If problem persists seek
technical assistance
Rechargeable batteries
Recharge batteries
completely flat
Problem/Cause
Battery charging in progress Wait
Batteries
completely
Wait for normal recharge cycle (24
flat, slow recharge cy-
hours max) or replace batteries
cle in progress
Unplug from power supply, check
Electrical problem with
that batteries are inserted cor-
circuit or batteries
rectly and restart recharge cycle.
If problem persists seek technical
assistance
Recharge cycle complete
Unplug from power supply
Faulty supply or internal
Seek technical assistance
technical problem
Device protection ac-
Disconnect and reconnect
tivated
power supply and start new
recharge cycle. If problem
persists seek technical as-
sistance
1
Solution