The disinfectant must be an electrolytic chlorine oxidiser (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Directions:
- Fill a container of suitable size to hold all of the parts to be disinfected with a solution of
potable water and disinfectant, according to the proportions indicated on the packaging
of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse with plenty of warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Sanitise the accessories 1A-1B-1C-2-3-4-5-6 by boiling them in water for 10
minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Sanitise the accessories 1A-1B-1C-2-3-4-5-6 with a hot steam steriliser for baby-
bottle (not the microwave type). Perform the process faithfully following the instructions of
the steriliser. To ensure that the disinfection is effective, choose a steriliser with an operating
cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND STERILISATION
Before disinfection or sterilisation, sanitise the nebuliser cup and the accessories, choosing
one of the methods described below.
method A: Sanitise accessories 1A-1B-1C-2-3-4-5-6 under potable hot water (approximately
40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories 1A-1B-1C-2-3-4-5-6 in the dishwasher with a hot cycle.
DISINFECTION
The disinfectable accessories are 1A-1B-1C-2-3-4-5-6.
The disinfectant must be an electrolytic chlorine oxidiser (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Directions:
- Fill a container of suitable size to hold all of the parts to be disinfected with a solution of
potable water and disinfectant, according to the proportions indicated on the packaging
of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse with plenty of warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
STERILISATION