Care Of Cooling Spray Nozzle Openings - Dentsply Sirona Midwest Stylus Plus Manual Del Operador

Ocultar thumbs Ver también para Midwest Stylus Plus:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

7 Reprocessing
52
7.3.3

Care of cooling spray nozzle openings

If your tap water is very hard, lime deposits may constrict or
completely block the cooling spray nozzles.
Interval
At least once a week
1.
Carefully clean the nozzle openings by running a cleaning wire
through them.
2.
Let the high-speed handpiece run briefly with cooling spray.
Operating instructions Midwest Stylus Plus / Midwest Tradition
67 27 148  D3673.201.09.01.02