7 Reprocessing
44
Instructions for Use Midwest Tradition / Tradition Pro / Quiet-Air
7.3.3
Care of cooling spray nozzle openings
If your tap water is very hard, lime deposits may constrict or
completely block the cooling spray nozzles.
Interval
●
At least once a week
1.
Carefully clean the nozzle openings by running a cleaning wire
through them.
2.
Let the high-speed handpiece run briefly with cooling spray.
67 27 155 D3673.201.10.02.02