6627587.book Seite 45 Dienstag, 14. November 2017 7:28 19
®
®
Operating Instructions Midwest Tradition
TB/Pro TBF / Midwest Tradition
TL/Pro TLF
7 Conditioning
7.4.3
Care of cooling spray nozzle openings
If your tap water is very hard, lime deposits may constrict or completely
block the cooling spray nozzles.
Interval
Interval
● At least once a week
1. Carefully clean the nozzle openings by running a cleaning wire
through them.
2. Let the high-speed handpiece run briefly with cooling spray.
45
66 27 587 D3673.201.02.01.02