Replacing The Spray Water Cartridge - Dentsply Sirona Midwest Stylus Plus Manual Del Operador

Ocultar thumbs Ver también para Midwest Stylus Plus:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

8 Maintenance
62
8.2.4
A
B
C
Operating instructions Midwest Stylus Plus / Midwest Tradition
Remove the LED and re-insert it after rotating it 180° around its
own axis.

Replacing the spray water cartridge

Required accessories
Spray water cartridge
Midwest Plus Spray
If the spray water cartridge leaks or is clogged.
1.
Remove the quick coupling from the supply hose.
2.
Remove the sealing washer (B).
3.
Click the control ring into the maximum water flow. The spray
water cartridge (A) can be replaced only in this position.
4.
Remove the spray water cartridge.
5.
Spray oil lightly onto the new spray water cartridge.
6.
Insert the new spray water cartridge.
67 27 148  D3673.201.09.01.02