Tecno-gaz B - TEST plus Manual Del Usuario página 26

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

 
 
INSTRUCTION OF THE USER 
 
1. 
Identify the biological indicator to be treated. 
THE  INSTRUCTIONS  ABOUT  HOW  TO  ACTIVATE,  TEMPERATURE  AND  DAUER  OF 
INCUBATION ARE INCLUDED IN THE USER' MANUAL OF THE BIOLOGICAL INDICATOR 
 2. 
Insert the biological indicator into the sterilizer. 
3. 
Perform the sterilization cycle. 
4. 
Remove the biological indicator from the sterilizer. 
5. 
Let the biological indicator cool for at least 10 minutes. 
6. 
Check the colour change to be sure that the sterilization has been properly carried 
out because the chemical indicator changes colour when it is subjected to steam. 
7. 
Break  the  glass  bulb  of  the  biological  indicator  which  is  inside  the  thermoplastic 
container.  Perform  this  operation  just  a  minute  before  inserting  the  indicator  in  the 
incubator. The indicator must be inserted in the incubator within and not beyond 1 hour 
from the moment it was removed from the sterilizer. 
8. 
Insert the biological indicator in the incubator's heating unit. 
9. 
Check the biological indicator at regular intervals to control the changing of colour. 
10. 
Incubate the biological indicator using as a reference an indicator not treated in a 
sterilisation cycle.  
11. 
For the detailed procedure carefully read the instructions for use which are inside 
the packaging biological indicators.  
WARNING:  "SEE‐TEST"  is  projected  to  incubate  biological  indicators  supplied  by  TECNO‐
GAZ.. 
TECNO‐GAZ will not be held responsible for the results obtained with other indicators.
 
11
 
ENGLISH
Tabla de contenido
loading

Este manual también es adecuado para:

2035ts

Tabla de contenido