Description Of Machine Operation - Tecno-gaz B - TEST plus Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

ENGLISH 
 
 
DESCRIPTION OF OPERATION 
The biological incubator is a device for use in medical and/or dental.
The producer is not responsible for damage to the unit and personal 
 
injuries caused by improper or lack of experience of operator.
Temperature of usage +17 / +27 ° C
 
On incubator there is an indicator light to indicate the possible bad‐functioning of 
the machinery or the momentary lack of electricity (and then the failure of tests in 
progress).  Also inside the device there is a thermostat security in case of reaching 
temperatures than those using interrupts the functioning of resistance. 
 
1.
Connect the power cord (fg.1 – POS.D) to the electrical line. 
2.
The warning led (fg.1 – POS.A) lights. 
3.
Push the START button (fg.1 – POS.E). The led switch off and the heating cycle 
of unit start. When occurs a drop of temperature, caused by natural cooling of 
the  unit,  the  electronic  card  order  to  switch  again  the  resistance.  In  case  of 
failure  to  reach  temperature  or  sudden  interruption  of  power  supply,  the  led 
stays on e the incubator stops working. 
 
 
Reliability of the validity of the test 
If at the end of incubation time (about 24 hours) the LED is off the test is valid. 
 
 
 If at the end of incubation time the LED is on, the test is NOT valid 
CAUTION:  HOT  SURFACE  –  Do  not  touch  the  heating  block  during 
operation of the unit.
 
CAUTION: Keep away from children 
 
 
CAUTION: Do not touch metal surfaces when the machine is running 
 
 
10
 
 
 
Tabla de contenido
loading

Este manual también es adecuado para:

2035ts

Tabla de contenido