Alarm History; Reviewing; Reviewing Patient Measurements; Deletingpatient Data - Riester RVS-100 Manual Del Usuario

Ocultar thumbs Ver también para RVS-100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 120
Note: After selecting [Alarm audio off], the icon will ap-
pear on the interface.
6.11 Reminder signal
When the active alarm audio is off, the alarm system would provide
a periodical audible reminder signal sound like "Ding, Ding, Ding".
[SETTINGS] > [ADVANCED], input the correct password to enter
the alarm control interface. In this interface, you can select or dese-
lect [Active reminder signal] to open or close the reminder signal.
You also can adjust the intervals between the reminder signal to be
30s, 60s, 90s, and 120s in this interface.
6.12 Resetting Alarm Limit
To reset all alarm limits to factory default levels, select [Alarm] >
[General] > [Reset alarm limits]. Limits will be reset to the fol-
lowing defaults:
%1B1=5D5B
<MDL
*%,
EE$?
,=<A9LJA;
*=GF9L9D
/H+

,.
0!),
Warning: A potential hazard can exist if different alarm
pre-sets are used for the same or similar equipment in
any single area.

6.13 Alarm History

Select the [ALARM] on the main measurement display and then
select the [HISTORY] tab to see the alarm time, alarm level, alarm
message, alarm duration time and so on of all the alarms as shown
in the next image:
Note:
The recorded number of the alarm logs depends on the
storage space.
The alarm system generates a technical alarm condition
when the storage space is insufficient. When storage is
less than 10MB, a low level technical alarm occurs, and
a prompt information will pop up as "Insufficient storage
space". When the storage space is less than 5MB, the
other low level technical alarm occurs, and a prompt infor-
mation will pop up as "Critical shortage of storage space".
When the alarm system is powered down, the log is
maintained, but the time of powering down will not be
captured in a log.
The contents of the log is maintained after the alarm sys-
55
*@@5B <9=9D
!?G5B <9=9D
/5/


%


),


/5/


%


),


/5/


%


),








tem has experienced a total loss of power (mains adapter
and internal electrical power source) for a finite duration.
When the log reaches capacity, the system will automati-
cally delete the earliest log.

7. Reviewing

You can use the Review feature to access any patient information
saved by the monitor.

7.1 Reviewing patient measurements

Select [REVIEW] on the home screen to access saved patient
measurement data.
7.2 Deleting patient data
Select the blank box
to the left of the Patient ID and then select
[DELETE] to delete the patient's measurement data.

7.3 Print patient data

Select the blank box
to the left of the Patient ID and then select
[PRINT] to print the selected patient's measurement data.

8. Battery

8.1 Introduction

The monitor can be fitted with a rechargeable battery to ensure
continuous operation in the event of a power outage. The battery re-
quires no special maintenance under normal conditions. While the
monitor is connected to an external power source, the battery will
charge, regardless of whether the device is turned on. In the case
of a sudden power outage, the monitor will automatically switch to
battery power without interruption of measurement.
Battery status can be found at the top right corner of the touch
screen.
indicates that the battery is fully charged.
indicates the battery is depleted and needs recharging.
indicates the battery is recharging.
indicates the battery is abnormal.
Battery power lasts for a limited time. When battery power is very
low, the monitor will issue a monitor technical alarm. The user
should immediately connect the device to a power supply to charge
the battery.
Caution: If the monitor is unlikely to be used for an ex-
tended time period, remove the battery prior to shipping
or storage.
Warnings:
Use only batteries specified in this manual.
Keep the batteries out of reach of children.
Check the battery regularly to guarantee its normal func-
tion.
Replace the battery at the end of its service life.
The battery can only be replaced, maintained by professi-
onal personnel specified by Rudolf Riester GmbH. Or the
device may not be started up.

8.2 Installing a Battery

The battery compartment is located on the bottom of the monitor.
Follow these steps when installing the battery.
1.
Turn off the monitor and disconnect the power cable and other
connected wires and cables.
2.
Open the battery door in the direction indicated on the door
label.
Tabla de contenido
loading

Tabla de contenido