Sensor Settings; Check Of Parameters; Preparations For Operation - Dräger X-am 2500 Instrucciones De Uso

Ocultar thumbs Ver también para X-am 2500:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 53
3.2

Sensor settings

The following parameters can be changed on the sensors:
Name
Alarm threshold A1 (in measuring unit)
Alarm threshold A2 (in measuring unit)
1
Interpretation type
Alarm threshold STEL
1
(in measuring unit)
Alarm threshold TWA
1
(in measuring unit)
1)
Interpretation only if the sensor is designed for this.
3.3

Check of parameters

To ensure that the values were correctly transferred to the gas detection
instrument:
1. Press the data from X-am 1/2/5x00 in Dräger CC Vision.
2. Check parameters.
Dräger X-am 2500
Range
0 - A2
A1 – measuring range end value
Inactive, TWA, STEL, TWA+STEL
0 – measuring range end value
0 – measuring range end value
4
Operation
4.1

Preparations for operation

WARNING
!
To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx.
See the marking on the rechargeable battery for permitted
rechargeable batteries and the corresponding temperature
class.
Substitution of components may impair intrinsic safety.
Before using the instrument for the first time, insert a charged
NiMH T4 power pack or batteries approved by Dräger, see
chapter 4.9.1 on page 28.
The instrument is now ready for operation.
Operation
23
Tabla de contenido
loading

Este manual también es adecuado para:

Mqg 0011

Tabla de contenido