NOTE!
Ensure containers are positioned properly in the respective holders.
NOTE!
Only use containers with overflow protection.
5.3.2
Connecting the Vacuum Tube
CAUTION!
When using the Aquilex System with tissue removal systems, e.g. MyoSure®, the
combination of low set pressures and excessive vacuum pressures may result in
a significant loss of intrauterine distension pressure which has the potential to
affect the visibility of the surgical field. Conversely, when employing a high dis-
tension pressure, the deactivation of the tissue removal system can lead to pres-
sure spikes that can exceed 150 mmHg.
Connect vacuum tube with hygiene filter to suction containers. The vacuum tube
with hygiene filter must be replaced when dirty and after 30 days at the latest.
The vacuum tube with hygiene filter should not be cleaned.
• Connection for low vacuum (white)
· Connect vacuum tube with white connector to low vacuum port (white) (1)
Fig. 5-5. This vacuum pump has a fixed vacuum pressure (~ 225 mmHg).
· Use the connecting tube ((5) Fig. 5-5) when two containers are serially con-
nected to the same vacuum port.
Aqu ilex
TM
Flui d
Ser.
Con trol
Nr. WO
Sys tem
100 -240
M
(1)
V / 50-6
YYY
0 Hz
/
Y-M M
Manu
W.O.
factu
rer
1058
M. WOR
(
7 Berlin
LD OF
MED
9633
+49
(0) 30
• Salzu
fer 8
ICINE
GER
7 Ludw
igssta
399
81-55
• GER
AG
MAN
Y •
(
dt • Alte
0
MAN
Y
+49
(0) 9263
Posts
877-0
traße
11
(2)
(3)
• Connection for high vacuum (green)
· Connect vacuum tube set with the green connectors to the high vacuum
port (green) (8) in Fig. 5-6. This vacuum can be adjusted to a maximum
500 mmHg using adjustment dial.
· Use the connecting tube ((12) Fig. 5-6) when two containers are serially
connected to the same vacuum port.
Fig. 5-5
Low vacuum tube
(1)
Connection for low vacuum
(white)
(5)
Hygiene filter
(2)
Vacuum tube
(3)
container
(4)
Connecting tube
(5)
(4)
System Operation
en
19