Formulae For The Preparation Of Solutions - Xylem SI Analytics TITRONIC 500 Manual De Instrucciones

Ocultar thumbs Ver también para SI Analytics TITRONIC 500:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Chapter 4 Method Parameters
89
After selecting a formula, please confirm your selection with <OK>/<ENTER>:
Fig. 58
The values of the individual parameters of the selected calculation formula can now be input one by one.
Fig. 59

4.5.2.2 Formulae for the Preparation of Solutions

A selection of special calculation formulae is available for the Prepare Solutions mode. The appropriate
calculation formula is selected on the Formula Selection submenu:
Fig. 60
A selection of 3 different calculation formulae is available:
W*(100-Fa-Fb)*Fc/Fd - W*(100-Fb) / (100*Fe) +Ff
W*(100-Fa-Fb)*(Fd/Fg ) - W*(100-Fb) / (100*Fg) +Ff
W*(100-Fa-Fb)*Fc / (100*Fd)