Commissioning, measuring, calibration

Smallest allowed bend radius
Commissioning
14
Connection cabel
Diameter
Plug type

Accuracy of the IDS measuring technique

Measured parameter
pH
U [mV]
T [°C]
Commissioning, measuring, calibration
Prepare the electrode for measuring as follows:
Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Open the refi lling opening for the reference electrolyte solution
(except BlueLine 32 Rx).
refi lling opening
The refi lling opening must always be open during calibration and
Measurement!
Note
Please keep the watering cap. It is required for the electrode to be stored.
Always keep the watering cap clean.
Remove any gas bubbles behind the pH membrane by shaking.
Connect the electrode to the meter.
Calibrate the electrode according to the operating manual of the meter and
observe the following rules while doing so:
4,3 mm
Fixed installation: 20 mm
Flexible use:60 mm
Socket, 4 pins
Accuracy (± 1 digit)
± 0.004
± 0,2
± 0.1
T 814252
BlueLine
05/2014