n reverse on load feature
for patient safety
The device automatically starts moving
in the opposite direction of the last
movement when the patient's resist-
ance (load) exceeds the set value.
Adjustable levels for reverse on load
feature: 1 – 25. At level 1, very low resist-
ance will cause the device to reverse; at
level 25, a high resistance is required to
initiate the reversal.
default setting: level 25
Caution!
m
Patient hazard –
The reverse on load feature is a safety
measure to protect the patient in the
event of cramps, spasms, locked joints
and similar situations. The manufac-
turer cannot be held liable for misuse of
this feature.
LEVEL 3:
n transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Set the
femur length on 49 cm and the lower
leg on 45 cm.
Select the function and press the START
key. The carriage moves to the transport
position. (see 6.3)
n new patient
With this function, the CPM device will
move to the home position, allowing the
mechanical settings to be completed.
Select the function and press the START
key. The device enters the home position
and existing therapy parameters will be
deleted.
With ARTROMOT®-K1 devices with
patient chip card, the factory defaults
will be restored. All values stored on the
chip card will be deleted.
The carriage will stop in the home
position.
ORMBS-12006_Artromot_K1_Comfort_2012_rz.indb 54
The "New Patient" function (home posi-
tion) selects the following settings:
− extension:
− flexion:
− speed:
− warm up:
− extension pause:
− lexion pause:
− timer:
− reverse on load:
− total therapy time:
n total therapy time
ARTROMOT®-K1 models
without patient chip card
The total therapy time is the added sum
of operating hours. If the device is used
by only one patient, this time is equiva-
lent to the duration of all the patient's
therapy sessions.
Under menu item "total therapy time"
of ARTROMOT®-K1 models with chip
card you can view each patient's total
therapy time (duration of all the pa-
tient's therapy sessions).
Deleting the stored therapy time Press
and hold the parameter key for 5 sec-
onds or select the New Patient function.
n Service MENU
For service purposes only, refer to
Service Manual.
Reminder:
You save the set parameter values
by pressing the STOP key.