6.
INSTALLATION: Triango 60 light head
DANGER
Danger to life through an electric shock.
When performing any work on the device, disconnect
it from the power supply or pull the connector out of
the socket and secure the device from being switched
on again.
WARNING
Risk of injury
The spring arm is under a high spring load. If the boom
is lowered without the device installed, it MUST be
held steady.
If released, it will spring up, which can cause serious
injury.
Only remove the body of the light when the boom
is in the up position or when a second person
is securely holding it in a lowered position.
Always have a second person to help when installing
or removing the body of the light, to avoid serious
injury or damage.
Connect the connector of the light head to the connector
of the spring arm.
Push the light head onto the spring arm.
When aligning the light head, make sure that the flat side
of the brass part points down to "A".
Screw the grub screw into the locking part "B"
The screw grips the previously aligned, flat side of the
brass part
36
A
B
WARNING
Risk of injury from falling light head.
Make sure the safety element is properly installed.
Material damage due to improperly installed light
head.
Tighten the screw as required by the friction of the
light head
Attach handle.
7.
OPERATION
Triango 60 C, W, F
WARNING
This product emits possibly hazardous optical
radiation. Do not stare at the light emitted from the
surgical luminaire. Eye injury may occur.
The optical radiation emitted by this product complies
with the exposure limits for reducing the risk of
photobiological hazards in IEC 60601-2-41.
Triango 60-1 C, W, F
1
Switch the light on or off with button "1"
Continuously dim the light by pressing and holding down
button "2".
2