SoftCloud™ Ace, SoftCloud™ Pro
3. Reconnect the connector(s) to disengage permanent
offloading. The air cell will be inflated again and return
to normal operation mode.
4. Close the push button flap A.
5.5 CPR Procedure
A
1. Firmly pull the yellow rapid release CPR tag A from the
side of the mattress to deflate the whole system.
2. Switch off the control unit.
Mattress will start to deflate.
3.
When CPR is complete replace the yellow CPR Tag
ensuring the sealing connectors are firmly attached and
restart the control unit following chapter: "Activating
the control unit". Using the target design to line up
each plug with it's corresponding socket.
20
The CPR tag can have three (3) of four (4) connectors
depending on mattress model.
IMPORTANT!
– Wait for the mattress system to gain optimal
pressure.
– Perform a Bottoming Out test after inflating
the mattress following rapid deflation.
5.6 Audible Warning functions
The audible warning LED flashes, and an audible warning
sounds, to indicate that the control unit or mattress pressure
has failed. The LED will remain illuminated until appropriate
pressure is restored. The audible warning can be silenced by
pressing the audible warning mute button.
The system has four different audible warning signals and
one service indicator, identified by five different symbols.
The symbols are illustrated below.
1630732-A