Usage
!
IMPORTANT
▸ To prevent accidental cover damage, do not place hypo-
dermic needles, venflons, scalpels or other similarly sharp
objects onto the mattress. Ensure that all venflons are taped
down correctly with no sharp edges exposed.
▸ When using bridging boards or other patient transfer aids,
care should be taken not to damage the mattress cover.
All transfer aids should be checked for any sharp edges or
burrs before use as these can damage the mattress cover.
▸ Make sure that the mattresses are not jammed or damaged
by sharp edges when used on beds with an adjustable frame.
▸ When using the Softform
bed ensure that the knee break is used before the backrest.
4.2 CPR Procedure
The Softform
Premier
Active 2 mattress has been fully tested to com-
®
ply with the current CPR standard of 4 – 5 cm compression depth. This
was achieved at all stages of inflation/deflation.
A report on these findings conducted by the Resuscitation Unit at the
University Hospital of Wales, Cardiff is available upon request.
▸ Hold down Red CPR button.
▸ Pull hose connector firmly away from the pump.
▸ Switch off the pump.
Mattress will start to deflate.
▸ When CPR is complete reactivate the system following chapter 3.2,
page
8.
Premier
Active 2 on a profiling
®
Softform
®
Premier Active 2 Mattresses
9