SoftCloud™ Ace, SoftCloud™ Pro
Emission test
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Electromagnetic immunity
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
34
Compliance
Electromagnetic environment – guidance
low-voltage power supply network that supplies buildings used for
domestic purposes.
Complies
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
± 1 kV line(s) to line(s)
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV differential mode
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
1630732-A