SoftCloud™ Ace, SoftCloud™ Pro
Recommended separation distances between portable and mobile RF communications equipment and the mattress system
Alternating Control Unit
The mattress system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the mattress system can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the mattress system as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
38
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.167√P
0.117
0.369
1.167
3.689
11.667
80 MHz to 800 MHz
d = 1.167√P
0.117
0.369
1.167
3.689
11.667
800 MHz to 2.5 GHz
d = 2.333√P
0.233
0.738
2.333
7.379
23.333
1630732-A