4.10 Preparing for CPR Procedure
For this procedure, refer to FIGURE 5.
1.
Press and hold the tabs on hose connector while pulling the hose from the control unit.
2.
Twist the CPR connector at the foot end of the overlay and all air will be evacuated in the torso and chest region of the overlay within 18 seconds.
4.11 About Power Outage and Transportation
If the hoses remain connected to the control unit, the overlay retains air during a power outage, during transportation, when the control unit is
unplugged or when the control unit is turned Off. If the hose becomes disconnected or damaged, the overlay deflates.
Part No 1176084
Hose Connector
Tab
FIGURE 5 Preparing for CPR Procedure
Hose
25
4 OPERATION
Foot-end of Overlay
RED CPR Connector
Twist Open/Twist Closed
Invacare® MA50CE Series