Display:
3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Calibration:
Automatically with up to three buffer solutions, sequence during calibration optional, freely
definable buffers can be input. Default buffer solutions according to DIN 19 266 and NBS, or
technical buffers: pH = 1.00; pH=4.00; pH=4.01; pH=6.87; pH= 7.00; pH= 9.18; pH=10.00;
Inputs:
Measurement input pH/mV:
Measurement input
Measurement input Pt 1000:
Power supply:
RS-232-C Interface:
Data bits:
Stop bit:
Start bit:
Parity:
Baud rate:
Address:
RS-232-1
RS-232-2
USB Interface:
USB –Typ B ("slave")
USB –Typ A ("master")
Stirrer connection:
Housing material:
Front keyboard:
Housing dimensions:
Weight:
Ambient conditions:
pH/mV-input with electrode socket according DIN 19 262/or
BNC
µA:
(Dead-Stop-) connector for double platinum electrode
Connection sockets: 2 x 4 mm)
Temperature sensor probe for P 1000 resistance thermometer
Pt 1000 (Connection sockets: 2 x 4 mm
power supply 90-240 V; 50/60 Hz, power input: 30 VA
Use the Power supply TZ 1853, Type No.: FW 7362M/12 only!
RS-232-C interface separated galvanically through photocoupler
Daisy Chain function available.
adjustable, 7 or 8 Bit (default: 8 Bit)
adjustable, 1 or 2 Bit (default: 1 Bit)
static 1 Bit
adjustable: even / odd / none
adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
adjustable, (0 to 15, default: 01)
for computer, input Daisy Chain
devices of SI Analytics, titrator TitroLine
TW alpha plus/TW 7400
- Burettes TITRONIC
- Balances of the types Mettler, Sartorius, Kern, Ohaus (for more, please contact SI
Analytics)
- Exit Daisy Chain
2 x USB-type A and 1 x USB-type B
for connecting a PC
for connecting:
- USB keyboard
- USB printer
- USB "mouse" ("mouse"),
- USB data media e.g. USB stick
- USB Hub
12V DC out, 500mA
power supply for stirrer TM 235, TM 135
Polypropylene
polyester coated
15.3 x 45 x 29.6 cm (W x H x D), height incl. interchangeable unit
ca. 2.3 kg for basic unit
ca. 3.5 kg for complete device incl. interchangeable unit (with empty reagent bottle)
Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °