pretitration volume can be set from 0.000 and 99.999 ml, the possible range for setting the waiting time following
pretitration is between 0 and 999 seconds.
Fig. 95
Titration end
The end of a titration is reached, and the result will be calculated as soon as, or if, respectively:
The defined End value pH, mV µA value has been reached
•
•
The criteria (steep, flat, slope value) have been met for one turning point (EQ1) or two turning points
(EQ2, only TitroLine
•
The predefined value ml has been reached (Maximum titration volume)
•
or if titration was terminated manually by operating the <Stop> key.
Fig. 96
It is also possible to switch off the criteria for the end value for pH and mV. This value cannot be switched off in
the case of a µA (Dead Stop) titration!
Fig. 97
The possible pH end value input ranges from 0.000 to 14.000.
The possible mV end value ranges from -2000 to +2000.
The range of the µA input can be selected between 0.0 and 100.0.
Automatic detection of the equivalence point (EQ) can be switched on and off for linear or dynamic titration.
Chapter 4 Method Parameters
®
7000) in the case of a linear or dynamic titration.