Adjusting The Knee Pad Height - Invacare Roze Manual Del Usario

Ocultar thumbs Ver también para Roze:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

NOTE: The actuator lead wires will be visible after the large base cover is removed.
6.
Remove the socket head screw and washer that attach the leg actuator to the base.
7.
At the other end of the actuator, remove the socket head screw and nut that attach the leg actuator to the leg bracket.
8.
If necessary, repeat STEPS 5‐7 on the opposite side of the base to remove the other leg actuator.
9.
To replace the actuator(s), reverse STEPS 5‐7.
Caster
Leg
Large Base Cover

Adjusting the Knee Pad Height

NOTE: For this procedure, refer to FIGURE 7.3.
NOTE: When adjusted properly, the patient's knee will make 
contact at the center of the knee pad.
1.
While supporting the knee pad, disengage the quick 
release knob on the side of the knee pad mounting 
bracket.
2.
Choose a height adjustment that will be comfortable for 
the patient and will provide the necessary support. 
Engage the quick release knob into position until the knee 
pad is secure in place.
Part No. 1164940
BOTTOM VIEW OF
PATIENT LIFT
Pan head screw
(eight screws attach the
large base cover)
Mast Handles
FIGURE 7.2 Replacing the Leg Actuators
27
SECTION 7—MAINTENANCE
(eight screws attach each
small base cover)
Small Base Cover
Knee Pad
Quick Release
Knob
FIGURE 7.3 Adjusting the Knee Pad Height
Roze™Stand Assist Patient Lift
Screw
Tabla de contenido
loading

Tabla de contenido