SECTION 6—TROUBLESHOOTING
SYMPTOMS
Noisy or dry sound from pivots.
Electric actuator fails to lift when button is
pressed.
Unusual noise from lift arm actuator.
Lift arms will not lower in uppermost position.
Legs don't open and close properly.
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.
Part No. 1164940
FAULTS
Needs lubrication.
Hand-control or actuator
connector loose.
Battery low.
RED emergency stop button
pressed IN.
Battery not connected properly
to control box.
The connecting terminals are
damaged.
Electric actuator in need of ser-
vice or load is too high.
Actuator is worn or damaged
or spindle is bent.
Lift arms require a minimum
weight load to lower from the
uppermost position.
Leg actuators may be worn or
damaged or spindle is bent.
23
SECTION 6—TROUBLESHOOTING
SOLUTION
Refer to Lubricating the Lift on page 25.
Charge batteries. Refer to Secondary Emer-
gency Release on page 14.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Secondary Emergency Release on
page 14.
Replace the battery pack. Refer to Second-
ary Emergency Release on page 14.
Refer to Replacing the Electric Actuator on
page 25.
Refer to Replacing the Electric Actuator on
page 25.
Pull down slightly on the lift arms.
Refer to Replacing the Leg Actuators on
page 26.
Roze™Stand Assist Patient Lift