Seca mBCA 555 Instrucciones De Uso página 179

Ocultar thumbs Ver también para mBCA 555:
Idiomas disponibles

Idiomas disponibles

Fault/error code
080-297XX-XXX
081-297XX-XXX
082-297XX-XXX
083-297XX-XXX
084-297XX-XXX
Cause
Volume set to zero
Loudspeaker is defective
Voice output memory cannot be read
Voice file not found
An error occurred during the measure-
ment
An error occurred during calibration
Interference caused by reflection
Interference caused by other ultrasonic
emitters
Ambient temperature too high or too
low
Temperature sensor is defective
Remedy
è Setting the vol-
Increase volume
ume, page 169
Inform seca Service
Inform seca Service
Deactivate beeps and voice output to sup-
press the error message until repair is car-
ried out:
• Deactivate announcement of patient in-
è Activating/deactivating an-
structions
nouncement of patient instructions (Mea-
surement), page 170
• Deactivate announcement of measured
è Activating/deactivating an-
results
nouncement of measured results (Re-
sults), page 171
è Activating/deactivat-
• Deactivate beeps
ing beeps, page 170
Inform seca Service
Deactivate beeps and voice output to sup-
press the error message until repair is car-
ried out:
• Deactivate announcement of patient in-
è Activating/deactivating an-
structions
nouncement of patient instructions (Mea-
surement), page 170
• Deactivate announcement of measured
è Activating/deactivating an-
results
nouncement of measured results (Re-
sults), page 171
è Activating/deactivat-
• Deactivate beeps
ing beeps, page 170
Repeat measurement and ask the patient
to keep still
If the error recurs, inform seca Service
• Remove objects from the immediate
vicinity of the device.
• Ask people in the vicinity to stay further
away from the device
Increase the distance from other ultrasonic
emitters
Observe ambient conditions for operation,
è General technical
transport, and storage
data, page 193
Inform seca Service
Troubleshooting • 179
English
loading