Measuring bioimpedance
User qualification for bioimpedance measurement
Performing a bioimpedance measurement
142 • seca 555/554/552; seca 550/549; seca 257/256
Device mode
Basic
Advanced
Expert
Service
Bioimpedance measurements may only be carried out by persons who have
been familiarized with the functions of the device according to the regulations of
the respective institution.
To perform a bioimpedance measurement, at least the following sections of
these instructions for use must be observed in addition to the information in this
section:
è Safety information, page 108
è Contraindications, page 106
• Hygiene treatment
WARNING!
Patient hazard due to infections
Diseases can be transmitted due to poor hygiene.
► Ensure that the patient does not have any infectious diseases.
► Make sure that the patient's hands and feet are clean.
► Make sure that the patient does not have any open wounds on the
palms of the hands or the soles of the feet.
► Disinfect the electrode surfaces after every measurement.
WARNING!
Injury from falling
► The device is not designed for supporting patients when getting
up, e.g. from a wheelchair. Assist people with limited motor skills
when they are getting up, e.g. from a wheelchair.
1. Ask the patient to step onto the weighing platform.
2. Determine the patient's weight and height as described in the relevant sec-
tions of these instructions for use.
ð The BIA start key is shown.
ð The foot electrodes for bioimpedance measurement light up on the
weighing platform.