5.4
Connecting a barcode scanner (optional)
126 • seca 555/554/552; seca 550/549; seca 257/256
NOTICE!
Faulty measurement due to reflections
If there are objects or people in the immediate vicinity of the device, in-
correct measurements will result.
► Ensure that there are no objects or people within 0.5 meters of the
front or side of the scale during the measurement procedure.
► Ensure that the device is at least 0.2 meters away from the wall.
► Ensure that the patient is not wearing any kind of hair accessory on
top of their head.
3. Align the device by turning the foot screws.
Foot screw
ð The device is positioned horizontally when the air bubble of the spirit level
is in the precise center of the circle.
CAUTION!
Injury from a lack of stability
If the foot screws are screwed out too far, they may come loose from
the device. The device will then be unstable.
► Screw the foot screws out a maximum of 10 mm.
► If the device cannot be aligned horizontally with the screws
screwed out as far as possible, the setup location is unsuitable. Set
the device up in a suitable location.
Knurled screw
4. Tighten the knurled wheels in the direction of the arrow.
ð The foot screws are secured against being adjusted.
A barcode scanner can be connected to the USB interface of the multifunctional
display.
The barcode scanner is required for the following functions:
Configuration
: Define network data in the
transmit them to the device using the QR code:
tions, page 161
: Record patient and user IDs for transmitting measured results to
Operation
the
seca analytics 125
measurement, page 145
Spirit
level
seca connect 103
è Setting up network func-
software or to an EMR System:
Air bubble
software and
è Completing the