APPENDIX: Guidance and manufacturer's declaration of conformity
(Table according to EN 60601-1-2)
Electromagnetic Emission
Guidance and manufacturer's declaration – electromagnetic emissions
The BEMER Horse-Set products are intended for use in the electromagnetic environment spec-
ified below. The customer or the user of the BEMER Horse-Set devices should ensure that they
are used in such an environment.
Emissi st
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Electromagnetic Immunity
Guidance and manufacturer's declaration – electromagnetic immunity
The BEMER Horse-Set products are intended for use in the electromagnetic environment specified
below. The customer or the user of the BEMER Horse-Set devices should ensure that they are used
in such an environment.
Immunity test
IEC 60601
standard
test level
Electrostatic
6 kV contact
discharge (ESD)
8 kV air
IEC 61000-4-2
Electrical fast
2 kV for power
transient / burst
supply lines 1 kV for
input/output lines
IEC 61000-4-4
Surge
1 kV line to line
IEC 61000-4-5
2 kV line to earth
Voltage dips, short
<5 % U
interruptions and
40 % U
voltage variations on
70 % U
power supply lines
<5 % U
IEC 61000-4-11
Compliance
Electromagnetic environment – guidance
Group 1
The BEMER Horse-Set devices use RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
Class B
The BEMER Horse-Set products are suitable for use in all establishments,
Class A
including domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Complies
Compliance
level
6 kV contact
8 kV air
2 kV for power supply
lines
1 kV line to line
2 kV line to earth
(0.5 cycle)
<5 % U
T
T
(5 cycles)
40 % U
T
T
(25 cycles)
70 % U
T
T
for 5 s
<5 % U
T
T
Electromagnetic environment –
guidance
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
(0.5 cycle)
Mains power quality should be that of a
typical commercial or hospital environment.
(5 cycles)
(25 cycles)
for 5 s
Note: U
is the a.c. mains voltage prior to application of the test level.
T
211