Problem
No powder flows out from
the polisher nozzle during
operation
No water flows out from
the polisher nozzle during
operation
44
Possible Cause
Dental unit not connected to
the air circuit
The air polisher container is
empty
Polisher nozzle clogged due
to an excessive amount of
moisture in the powder or
insufficient
cleaning/maintenance
air polisher device channel
clogged due to an excessive
amount of moisture in the
powder or insufficient clean-
ing/maintenance
The amount of powder in
the container exceeds the
maximum level
Dental unit not connected to
the water circuit
Too much limescale in the
water
Dental unit water filter
clogged
Solution
Check the connection
between the dental unit and
the air circuit
Fill the powder container
without exceeding the maxi-
mum level (read
Chapter 5.2
on page 14
, point 9)
Read
Chapter 5.3 on page
18
,
Chapter 6.1 on page
22
Chapter 6.5 on page
,
26
and
Chapter 7 on page
38
.
Remove the powder from the
container and clean it using
a dry cloth
Read
Chapter 7 on page
38
Restore the correct level of
powder in the container (see
Chapter 5.2 on page 14
,
point 9)
Check the connection
between the dental unit and
the water circuit
Immerse the polisher nozzle
in a 2% acetic acid solution
for a few minutes. For the
complete cleaning procedure
see
Chapter 6.5 on page
26
Check the connection to the
dental unit water circuit
loading