CLEANING, DISINFECTION AND STERILISATION
Sequence and parameters applicable to
the cycle:
• 1 min, rinse with cold water;
• 5 min, wash with alkaline detergent
at 55 °C ±2 °C;
• 1 min, neutralisation with
neutralising liquid, 1/3 cold water
and 2/3 hot water;
• 1 min, rinse with water (1/3 cold
water, 2/3 hot water);
• 5 min, thermal disinfection at 93 °C
with demineralised water.
Thermal disinfection has not been tested
experimentally. In compliance with ISO
15883-1, Table B.1 [4] thermal disinfec-
tion at a temperature of 90 °C for 5 min-
utes determines a value of A0 3000.
6.6
Cleaning Control
6.6.1
Required Materials
• Light source;
6.6.2
Procedure for Cleaning Control
Once the cleaning operations have been
completed, check the polisher nozzles,
K9 wrench, and cleaning needle under
an adequate source of light, if necessary
using a magnifying glass 2.5X, paying
attention to any parts that may conceal
residual soil (threading, cavities, grooves)
and, if need be, repeat the cleaning cy-
cle if soil is still visible. Finally, check the
integrity of any parts and elements that
may have deteriorated during use.
2
• 2.5x Magnifier.
1
+
_ _
-