CLEANING, DISINFECTION AND STERILISATION
6.5.1
Manual Cleaning
Manual cleaning can be carried out as an alternative to the automatic cleaning described in
Chapter 6.5.2 on page 34.
6.5.1.1
Required Materials
• Enzymatic detergent at pH 6-9;
• Water;
• Container for immersion in the
enzymatic liquid;
• Ultrasonic tank;
Obtain a bucket and ultrasonic tank large
enough to hold the accessories needing
to be cleaned.
Prepare an enzymatic detergent solutio-
n
at pH 6-9 according to the manufac-
a)
turer's instructions, in a sufficient quan-
tity to fill the bucket and ultrasonic tank.
Use warm water for the bucket (30 °C - 35
°C) and water at ambient temperature (22
°C ±2 °C) for the ultrasonic tank.
CAUTION: Once used, dispose of the
enzymatic detergent correctly, do not
recycle it.
a) Process validated by independent bodies with
Mectron ENZYMEC enzymatic detergent, 0.8%
v/v.
• Clean, soft, low-lint cloths;
• Brush with soft nylon bristles;
• 25 ml syringe;
• Demineralised water.