- use the
- press the
- use the
- press the
LAS on
to enable the laser device
or LAS oFF
to disable the laser device
- press the
Caution
If the laser indicator is enabled, the application of the adhesive weights (except for the ALU1
P and ALU2 P balancing programs with the weight-holder device present) and the refer-
ence for selecting the two weights behind the spokes (Hidden Weight) is at 6 o'clock (H.6).
16. UNBALANCE CALIBRATION
WITH CALIBRATION WEIGHT
This programme allows sensitivity calibration whenever calibration when it appears to be
out of the tolerance range or when the machine requires this procedure to be performed
by displaying the E 1 error message.
Proceed as follows to recall this programme:
• press the
• use the
• press the
1. Fit a wheel of average dimensions (diameter no less than 14") on the wheel balancer
with preferably only a small unbalance.
2. The machine will display the following message:
• CAL on the left-hand display;
• GO on the right-hand display.
3. Perform a first spin.
4. At the end of the spin, fix the calibration weight supplied with the machine on the
136
keys to search for SEt UP
key to confirm
keys to search for LAS on (LAS oFF if previously set)
key to change the setting:
key to confirm and exit.
key
keys to search for Unb CAL
key to perform calibration, proceeding as follows:
EM 9250 - EM 9250 C CompactLine Operator's manual