A
B
C
D1
Door
closed
D2
Door
opened
E
F
G
H
Minimum
distance to
maintain
1. Install the autoclave in an environment suitable for carrying out sterilization procedures.
2. The room must be adequately lit and ventilated, as provided for in the directives in
force.
3. Install the autoclave away from heat sources and water splashes.
4. Position the autoclave on a surface that is able to bear its weight. The minimum
recommended load bearing capacity of the surface is 80 kg.
5. The supporting surface must be perfectly horizontal and must not have inclinations.

INSTALLATION

FIG.23
EUROPA B EVO
525 mm
510 mm
400 mm
370 mm
426 mm
455 mm
440 mm
EUROPA B EVO 24
666 mm
75 mm
410 mm
515 mm
50 mm