Default intensity: sets the intensity of the stimulus from which to start
the frequency change for the manual examination. Default value: 40 dB
HL.
-
Automatic frequency skip function: enables / disables automatic
switching to the next frequency after a value is saved. Default value:
disabled.
-
Switch mode: allows the interrupter button to be used as a button
(stimulation active as long as the key is pressed) or switch (stimulation
is activated when the key is pressed and deactivated when it is pressed
again). Default value: pulsating.
-
VA transducer: sets the type of transducer for air conduction, allowing
the user to choose between supra-aural and insert earphones. Default
value: supra-aural headphones.
7.2.7
Display brightness
The brightness of the display can be set between 20% and 100%. Default value:
80%.
7.2.8
Printer
Access to the print options menu:
-
Print patient details: allows enabling the printing of the patient's
personal details. Default value: enabled.
-
Print reflex graphs: allows printing the acoustic reflexes in graph form.
Default value: disabled.
7.3
L
ICENSE MENU
Access the menu to enable additional licenses.
EN
29