Laktate Lak-EN310 Manual De Instrucciones página 52

Analizador de glucosa y lactato
Idiomas disponibles

Idiomas disponibles

Perform quality control testing
Illustration
52 | Blood Glucose And Lactate Meter User's Manual
Steps
Step 1: Press the power button to
enter the power-on interface.
Step 3: Remove a blood glucose or
blood lactate test strip from the con-
tainer and immediately seal the cap.
Step 5: At this point, the meter is
turned on, and a flashing blood drop
icon appears on the screen, indicating
that the meter is ready.
Illustration
Front end of the
test strip slot
Dividing line of a
test strip
Steps
Step 2: Press the down arrow key on
the side to enter QC mode. The top of
the display shows "QC".
Step 4: Insert the test strip with the
logo side up into the test strip slot
and push it all the way in until you
hear "beep".
Step 6: Shake the control solution
vial to mix the control solutions. After
discarding the first drop of control
solution, wipe the dropper tip. When
the control solution touches the bot-
tom side of the test strip, it will be au-
tomatically siphoned in. The reaction
zone is filled with control solution
when you hear a "beep". Then the
hourglass icon will start flashing.