function to add a new calibration point. As recommended,
measure one or more times on the calibration pad (provided).
The "Source" value shown on the display is the current
measurement result, while "Target" is the expected standard
value. The user can change the target value to the standard
value. After calibration, press the button labeled "Save" to save
the calibration settings and return to the previous menu.
IMPORTANT: if a zero calibration is performed at that time, the
first value in the point calibration will be "0". This value cannot
be edited or deleted, so in this case you must select "Edit" to
calibrate the calibration point.
IMPORTANT: before point calibration you will be advised to set
the Probe mode to Auto mode. After point calibration, zero
calibration is not recommended.
Then after calibration, "Clear Magnetic" will appear to clear the
calibration data based on the magnetic properties of the
substrate and "Clear Eddy" will appear to clear the calibration
data based on the chip currents.
- "Clear all" - is a function to clear all calibration settings, both
zero and spot.
• device settings
With the device turned on, press the button assigned under the
"Menu" function to enter the menu mode.
Select "Option" item to go to the interface of changing device
settings. Selecting an item, press the button assigned as "Enter"
to go to the setting of the parameter:
CAUTION!
- Probe Mode" - "Eddy Current Mode" (eddy current mode
used only for measurements of coatings on non-
ferromagnetic surfaces - "NFe" icon will appear during
measurement), "Magnetic Mode" (magnetic field mode used
only for measurements of coatings on ferromagnetic
surfaces - "Fe" icon will appear during measurement) or
"Auto Mode" (automatic selection of measurement method
by the device and determination of substrate type).
- "device" - select μm, mm, mile or inch.
denotes the currently selected item
means that the function is active
loading