15.3 Use of the microscope (B-510LD series)

The filter turret is provided with 4 positions.
In each of the four positions a fluorescence filter can be in-
serted, which can be selected from the options shown in the
table below.
You can always add an additional filter after the first in-
stallation (see section 8.6).
If all four turret positions are full, the observation in
transmitted light will be affected by the presence of the
fluorescence filter.
1. Turn on the main switch ①. (Fig. 96)
For the model B-510LD4D: turn on the microscope using
the main switch on the right side of the frame. (Fig. 53)
2. Move the filter selector ② to one of the available positions
until the click stop. (Fig. 97).
When the filter is in the correct position, the dedicated LED
lights up.
3. Adjust to the desired brightness by rotating the dial ①.
When switching the fluorescence filter, the LED light
goes out. This is not a defect.
F I L T E R
EXCITATION
NAME
FILTER
M-1223
325-375 nm
M-1223.1
340 - 390 nm
M-1222
390-420 nm
M-1220
455-495 nm
M-1220.1
455-495 nm
M-1221
510-550 nm
M-1221.1
510-550 nm
M-1228
582-603 nm
M-1224
590-650 nm
M-1225
595-645 nm
M-1226
623-678 nm
M-1227
720-760 nm
D I C H R O I C
BARRIER
MIRROR
FILTER
415 nm
435LP nm
405 nm
420 - 470 nm
440 nm
450LP nm
500 nm
510LP nm
500 nm
518 - 542 nm
570 nm
575LP nm
570 nm
585 - 625 nm
610 nm
615-645 nm
660 nm
665LP nm
655 nm
665-715 nm
685 nm
690-750 nm
770 nm
780LP nm
DAPI
Hoechst
DAPI
Hoechst
Coumarin
Pacific Blue
FITC
GFP
FITC
GFP
Rhodamine
TRITC
Propidium Iodide
Rhodamine
TRITC
Propidium Iodide
Red#2 FISH
Cy5
Cy5
AlexaFluor 647
Draq5
Cy5.5
Li_Cor IR dye 800