Installing Fluorescence Filter (B-510Ld Series) - Optika B-510 Serie Manual De Instrucciones

Ocultar thumbs Ver también para B-510 Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122
8.9

Installing fluorescence filter (B-510LD series)

1. Disconnect the power supply plug from the fluorescence il-
luminator.
2. Open the side cover of the illuminator, by unscrewing the
side screws ①. (Fig. 43)
It could be helpful to remove the observation head.
The cubes are mounted on the opposite side of the cover:
opening the left cover acts on the right side of the slider and
vice versa.
3. Open the top door of the fluorescence illuminator by un-
screwing the four screws ② and release the cover. (Fig. 44)
4. Loosen the front locking screw ③ on the fluorescence cube
slider, corresponding to the fluorescence cube to be installed
or replaced. (Fig. 45)
5. Insert the fluorescence cube into the dovetail ④ of the cube
slider and move it to the click position. (Fig. 46)
6. Tighten the locking screw ③. (Fig. 45)
Page 35
F ig. 43
F
ig. 43
F ig. 44
F
ig. 44
F F ig. 45
ig. 45
F ig. 46
F
ig. 46
Tabla de contenido
loading

Tabla de contenido