10. Use of the microscope (B-510BF/B-510ERGO/B-510-2-2F-3-5)
10.1 Light intensity adjustment
Operate on the light intensity adjustment knob to turn ON/OFF
the microscope and to increase/decrease the illumination voltage
. (Fig. 52)
①
•
For B-510LD4D only: The microscope has a main switch
located on the right side of the stand. (Fig. 53)
10.2 Coarse focus tension adjustment
•
Adjust the tension using the provided tool.
The coarse knob tension is pre-set in the factory.
To modify the tension according to personal's needs, rotate the
ring ② using the provided tool. (Fig. 54)
•
Clockwise rotation increases the tension.
If the tension is too loose, the stage could go lower by itself or
the focus easily lost after fine adjustment. In this case, rotate the
knob in order to increase the tension.
10.3 Coarse upper limit lever
The upper limit knob has two functions: prevent the contact
between slide and objective and acts as "focus memory".
1. After focusing the specimen, pull the lever ③ toward the
front of the microscope and lock it. (Fig. 55).
•
In this way the focus upper limit is set.
2. Now one can lower the stage with coarse focus knob, replace
the specimen and raise again the stage up to the upper limit:
specimen will be in approximate focus and will need a fine
adjustment to get the proper focus.
•
Fine focus movement is not affected by the coarse
focus lock.
•
To unlock, move the lever in the opposite direction to the
one used for the locking.
③
Page 39