5
Place the device on a firm, flat surface near your bed. Make sure that it is placed
at a level lower than your sleeping position.
CAUTION
!
Be careful not to place the device where it can be bumped into and knocked
over.
6
WARNING
!
Do not fill the water chamber while it is attached to the docking station and
the CPAP, APAP, or bilevel device.
Turn the water chamber so that the air outlet points up. Over a sink, carefully fill
the water chamber through the air outlet up to the maximum water level mark
shown on the side of the water chamber base.
The HumidAire 2iC should be filled with fresh water before use. Always ensure
that the water chamber is clean. See "Cleaning and Maintenance" on page 15.
WARNING
!
Do not fill the water chamber with hot or chilled water.
Do not overfill the water chamber, as water may enter the CPAP , APAP , or
bilevel device and air tubing.
CAUTION
!
Do not use additives, in particular aromatic additives (eg, scented oils and
perfumes). They may reduce the humidification output of the device and/or
cause deterioration of the water chamber materials.
Air outlet
Maximum
water level
mark
Preparing For Use
11