8/16 or 4 to 20 mA model
Test button
Sensor switches at high level
Fail-safe function will default to full signal
Sensor switches at low level
Fail-safe function will default to empty signal
Vibration amplitude diagnosis
NAMUR module (IEC 60947-5-6) and and 8/16 mA or 4 to 20 mA model
Measurement quality is related to the vibration amplitude of the fork. The diagnostic LED
indicates the quality of the vibration being sent to the LVS200 electronics.
Diagnostic LED off: measurement quality is good. The vibration amplitude is strong.
Diagnostic LED blinking: measurement quality is poor and vibration amplitude is
decreasing as fork becomes encrusted. When this happens, set the sensitivity
switch to decreased sensitivity.
Diagnostic LED on: vibration has stopped and fork is fully encrusted with material.
Current output setting
8/16 mA
The chart below illustrates the output current when:
Fork is clean
Fork is encrusted: weak vibration amplitude is shown
Fork is fully encrusted and vibration has stopped.
D Off
Diagnosis
I = 16 mA
Setting
D On
Signal
output LED
Diagnosis
LED
7ML19985FT62
SITRANS LVS200 – INSTRUCTION MANUAL
All manuals and user guides at all-guides.com
I = 16 mA
I = 8 mA
I = 20 mA
diagnostic LED
sensitivity adjustment:
B = factory setting
A = decreased sensitivity
filling or emptying
output setting
signal output LED
I = 8 mA
I = 8 mA
I = 16 mA
I = 6 mA
Page EN-15