Thermocouple (TC) connections
Attach the TC wires to the applicable TC mini-connector
(Figure 9). The wider blade is the negative. Then attach the
connector to the instrument.
Measure/simulate a Thermocouple
To measure/simulate the TC values:
1. Connect the instrument (Figure 9) and, if necessary,
adjust the Set Up (Table 1).
2. Select the task from the task selection menu (Table 2).
3. Select Settings ( ) to change the operation from
Temperature to mV or mV to Temperature.
4. If necessary, adjust the Settings (Table 4/5) and/or the
output values to the system (Table 7).
a) °C or °F (Input)
Edit
c) °C or °F (Output)
Figure 9: Example configuration - To measure/simulate the
temperature (°C/°F) or mV values of a TC
Transmitter calibration
To calibrate a transmitter:
1. Connect the instrument (Figure 10/11) and, if necessary,
adjust the Set Up (Table 1).
8
- [EN] English
K
°C
Settings
Settings
Settings
b) mV (Input)
Settings
Settings
Edit
Settings
d) mV (Output)
2. Select the applicable calibration task from the task
selection menu (Table 2/3) and, if necessary, adjust the
Settings (Table 4/5).
3. Supply the output values to the system (Table 7).
Figure 10: Example configuration - Transmitter calibration
with external loop power
Figure 11: Example configuration - Transmitter calibration with
internal loop power
Switch test
To do tests on a switch:
1. Connect the instrument (Figure 12) and, if necessary,
adjust the Set Up (Table 1).
2. Select the applicable switch test from the task selection
menu (Table 2/3) and, if necessary, adjust the Settings
(Table 5). The display shows the switch condition (open
or closed) in the top right-hand corner.
3. Supply the output values to the system (Table 7).
• Example - "Nudge" output.
Maximum: 30V
Edit
Settings
Edit
Settings