Note: Use the Dual Function area (
the same time. If the area
this area to off (). This saves the battery power.
Table 2: Menu options - Task selections (Area
Options
Description
(If applicable)
Input measurement tasks:
mA Measure ±55 mA
V Measure ±30V
mV Measure ±120mV
Hz Measure the frequency (Units: Table 4)
RTD Measure RTD temperature
 Measure RTD resistance or 
TC Measure thermocouple temperature OR mV
Only when an IDOS UMM is attached. An IDOS
measurement task.
Output tasks:
mA Supply 0 to 24 mA
V Supply 0 to 12V
mV Supply 0 to 120mV
Hz Supply an output frequency (Units: Table 4)
RTD Simulate RTD temperature
 Simulate RTD resistance or 
TC Simulate thermocouple temperature OR mV
Table 3: Menu options - Task selections
(Dual Function, area
Options
Description
(If applicable)
White button = A Dual Function is set.
Black button = Dual Function, area
Input measurement tasks:
mA Measure ±55 mA
V Measure ±30V
mA/24V Measure ±55 mA (24V loop power is on)
A switch test
Only when an IDOS UMM is attached. An IDOS
measurement task.

Set up the settings

When the task is set up (Table 2/3), use the Settings menu to
adjust the input and/or output operation.
Display: Task
Settings selection
1
mA + IDOS
 
Sk1 = Start/Stop
Sk2 = Settings
Menu:
4A
Settings
... Units
(Table 4)
4
- [EN] English
 ) to do two operations at
 selection is not necessary, set
 )
is set to off.
2
3
(If applicable)
Display:
5A
6A
... units
bar
Pa
(Table 4)
Menu:
Settings
 )
(Table 5)
If there is additional data for a menu option, select
Settings ( ) to see the values that are set up. If necessary,
adjust the values.
Table 4: Menu options - Settings (Input)
Options
(If applicable)
... Units Pressure Units (UPM only). If you select an IDOS task
...
CJ ... (TC only). To select the type of cold junction (CJ)
... type Select RTD Type (RTD only). To select an applicable
Trigger level
Table 5: (Part of table) Menu options - Settings (Output)
Options
(If applicable)
... Units Pressure/Temperature: Refer to Table 4.
7A
...
CJ ... (TC only). Refer to Table 4.
... type Refer to Table 4.
[]
Amplitude
4B
5B
[]
Description
(Table 2/3). Select one of the fixed units of
measurement (psi, mbar ... ).
Temperature Units (RTD or TC only). To select the
temperature units (°C or °F).
Frequency Units (Hz only). To select one of these
units:
Hz: Range  1000Hz
kHz: Range 0 to 50kHz
counts/minute (cpm)
counts/hour (cph)
(TC only). Change the measurement operation:
Temperature to mV OR
mV to Temperature
compensation.
Automatic: The instrument monitors the CJ
temperature and applies the necessary CJ
compensation.
Manual: Measure the CJ temperature and set the
applicable value. The instrument uses this value to
apply the necessary CJ compensation.
RTD type (Pt50, Pt100 ... )
Select TC Type (TC only). To select an applicable
thermocouple type (K, J, T ... )
(Hz only). To set the amplitude at which the instrument
senses a frequency signal. Default = 5V.
Auto Detect []/[ ]: Set this option to make the
instrument calculate the value from the available
signal.
(UPM only). Gage sensors or sensors with differential
operation. A zero correction that makes the instrument
read zero at local pressure.
(Leak Test only). To set an applicable period for the
leak test (Hours:Minutes:Seconds).
Description
Frequency Units (Hz only). To select one of these
units:
Hz: Range  1000Hz
kHz: Range 0 to 50kHz
pulses/minute (ppm)
pulses/hour (pph)
(TC only). Change the output operation:
Temperature to mV OR
mV to Temperature
(Hz only). To set the amplitude of the output signal.
Amplitude = 5V (Default).
To select and set up a value for the "Nudge" output.
Example: 1.000 mA increments.
Additional data: Select Settings ( )
K0405 Revision B